CYS3 Antibody

Shipped with Ice Packs
In Stock

Description

Structure and Biochemical Properties

Cy3-conjugated antibodies consist of an immunoglobulin (IgG) backbone chemically linked to the Cy3 dye. The conjugation process involves NHS ester chemistry, ensuring stable covalent bonds between the fluorophore and antibody lysine residues . Key characteristics include:

  • Excitation/Emission: Peak excitation at 550 nm and emission at 570 nm, compatible with standard TRITC filter sets .

  • Brightness: Higher extinction coefficient (ε=150,000M1cm1\varepsilon = 150,000 \, \text{M}^{-1}\text{cm}^{-1}) and quantum yield (ϕ=0.15\phi = 0.15) compared to TRITC, reducing background noise .

  • Stability: Sulfonated Cy3 resists aggregation in aqueous and non-polar media, enhancing performance in dehydration-heavy protocols .

Table 1: Cy3 vs. Common Fluorescent Dyes

ParameterCy3FITCTRITC
Excitation (nm)550495541
Emission (nm)570519565
PhotostabilityHighModerateLow
Compatible FiltersTRITCFITCTRITC
Common ApplicationsIF, FCM, IHCIF, ELISAIF

Immunofluorescence (IF) and IHC

Cy3-conjugated secondary antibodies (e.g., Goat Anti-Rabbit IgG) enable multiplex staining. For example, Cy3-labeled antibodies detected GFAP in rat brain sections with high specificity .

Flow Cytometry (FCM)

These antibodies facilitate rapid, quantitative cell surface marker analysis. A 1:20–1:100 dilution range is typical for CD206+ macrophage detection in arthritis models .

Antibody Microarrays

Full Moon BioSystems’ arrays use Cy3-labeled antibodies as positive controls, providing boundary markers for high-throughput protein profiling .

Sensitivity and Specificity

  • Signal-to-Noise Ratio: Cy3’s brightness allows detection at low antibody concentrations (e.g., 0.1 µg/mL in IHC) .

  • Cross-Reactivity: Anti-human IgG (Fc-specific) Cy3 conjugates show no cross-reactivity with mouse or rat serum proteins .

Table 2: Validation Metrics for Cy3-Conjugated Antibodies

ParameterValue/OutcomeSource
Dilution Range (IF)1:20–1:800
Stability (Storage)1 year at -20°C (50% glycerol)
Photobleaching Resistance>6 hours under continuous imaging
Multiplex CompatibilityCompatible with Alexa Fluor 647

In Vivo Macrophage Tracking

Cy3-tilmanocept, a CD206-targeting probe, identified inflammatory macrophages in murine arthritis models. Ex vivo imaging revealed Cy3 fluorescence colocalized with CD206+ cells in synovial fluid of rheumatoid arthritis (RA) patients, outperforming osteoarthritis controls .

Pathway-Specific Protein Profiling

Cy3-streptavidin detection in antibody microarrays enabled simultaneous analysis of 400+ proteins in RA synovial tissue, highlighting upregulated inflammatory markers (e.g., TNF-α, IL-6) .

Advantages Over Alternative Fluorophores

  • Brightness: 2–3× higher signal intensity than TRITC .

  • Versatility: Compatible with argon (514 nm) and helium-neon (543 nm) lasers .

  • Low Background: Minimal nonspecific binding in FFPE tissues .

Product Specs

Buffer
Preservative: 0.03% Proclin 300
Constituents: 50% Glycerol, 0.01M Phosphate Buffered Saline (PBS), pH 7.4
Form
Liquid
Lead Time
Made-to-order (14-16 weeks)
Synonyms
CYS3 antibody; FL3-27 antibody; At2g40880 antibody; T20B5.8Cysteine proteinase inhibitor 3 antibody; AtCYS-3 antibody
Target Names
CYS3
Uniprot No.

Target Background

Function
CYS3 Antibody is a specific inhibitor of cysteine proteinases. It is likely involved in the regulation of endogenous cellular processes and in the defense against pests and pathogens.
Database Links

KEGG: ath:AT2G40880

STRING: 3702.AT2G40880.1

UniGene: At.19895

Protein Families
Cystatin family, Phytocystatin subfamily
Subcellular Location
Secreted.

Q&A

What is the difference between Cy3 dye and anti-Cy3 antibodies?

Cy3 is a fluorescent cyanine dye commonly used to label proteins in proteomics research, particularly in antibody microarray applications. The dye forms covalent bonds with proteins through amine groups. In contrast, anti-Cy3 antibodies are immunoglobulins specifically developed to recognize and bind to the Cy3 molecule itself. These antibodies are utilized in various applications including ELISA, immunocytochemistry (ICC), immunoprecipitation (IP), dot blotting (DB), and Western blotting (WB) . The distinction is important as researchers may use either Cy3-labeled antibodies (antibodies tagged with Cy3 dye) or anti-Cy3 antibodies (antibodies that target the Cy3 molecule) depending on their experimental design.

What are the primary research applications for anti-Cy3 antibodies?

Anti-Cy3 antibodies serve multiple functions in research settings. They are primarily employed in applications such as ELISA, immunocytochemistry, immunofluorescence (IF), immunoprecipitation, and dot blotting . In microarray experiments, these antibodies may be used for quality control or signal amplification. Commercial anti-Cy3 antibodies are available in both monoclonal (e.g., clone EA08 from Creative Diagnostics and clone CY96 from Abcam) and polyclonal formats (e.g., from Creative Diagnostics), offering researchers flexibility based on their specific experimental requirements . Each application requires appropriate validation to ensure antibody specificity and performance in the intended experimental context.

How do Cy3 and Cy5 function in proteomics research?

Cy3 and Cy5 are fluorescent dyes extensively utilized in antibody microarray techniques for differential protein expression analysis. These dyes serve as markers that enable detection and quantification of proteins bound to antibody microarrays . Typically, two different protein samples are labeled with Cy3 and Cy5 respectively, allowing for comparative analysis on the same microarray. This approach offers significant advantages over cDNA microarray analysis as it directly measures protein expression levels rather than mRNA, addressing the often poor correlation between transcriptomic and proteomic data . The spectral properties of these dyes allow for simultaneous detection with minimal overlap when appropriate scanning parameters and filters are employed. The coupling of these cyanine dyes to proteins occurs universally through amine groups, making them versatile tools for protein labeling across various proteomics platforms .

How should researchers optimize protein-to-dye ratios when labeling with Cy3/Cy5?

Optimizing protein-to-dye ratios is critical for achieving reliable results in microarray experiments. Based on published methodologies, researchers typically use approximately 33.334 μg of labeled proteins in their experimental designs . The procedure involves adding 100 μL of protein sample to 100 μL of labeling buffer, followed by the addition of fluorescent dye. After incubation, excess unbound dye is removed via gel filtration using desalting columns .

The optimal ratio should provide sufficient signal intensity without causing excessive protein modification that might alter binding properties. While specific ratios may vary depending on the particular experiment, validation studies have demonstrated reliable results with ratios of 2:1 when comparing different fluorophore-labeled samples (e.g., 33.334 μg of one dye-labeled protein mixed with 16.667 μg of the alternative dye-labeled protein) . Researchers should perform preliminary experiments to determine the optimal ratio for their specific protein samples and detection system.

What controls should be incorporated when using anti-Cy3 antibodies?

When designing experiments with anti-Cy3 antibodies, multiple controls should be incorporated:

  • Specificity controls: Include both positive controls (Cy3-labeled proteins) and negative controls (unlabeled proteins) to confirm antibody specificity. Commercial microarrays often include fluorescence-labeled albumin as positive controls and non-labeled albumin as negative controls .

  • Cross-reactivity assessment: Test the anti-Cy3 antibody against similar fluorophores (particularly Cy5) to evaluate potential cross-reactivity.

  • Dye-swap experiments: For differential expression studies, perform reciprocal labeling of samples with Cy3 and Cy5 to identify and correct for dye-specific biases .

  • Internal technical replicates: Utilize duplicate spots on microarrays (as implemented in commercial systems like ClontechTM antibody microarrays) to assess technical reproducibility .

These controls collectively enhance data reliability and facilitate accurate interpretation of experimental results.

How can researchers assess the quality and specificity of commercial anti-Cy3 antibodies?

Evaluating commercial anti-Cy3 antibodies requires a systematic approach:

  • Literature validation: Prioritize antibodies with citations in peer-reviewed literature. For example, certain commercial options like Abcam's Anti-Cy3/Cy5 antibody [CY96] have multiple citations supporting their application .

  • Specificity testing: Perform controlled experiments using Cy3-labeled and unlabeled proteins to confirm specific binding.

  • Cross-reactivity assessment: Test the antibody against other fluorophores, particularly structurally related dyes like Cy5.

  • Linear response validation: Verify that the antibody produces a linear signal across a range of Cy3 concentrations, essential for accurate quantification.

  • Application-specific validation: Confirm performance in your specific application (e.g., ELISA, ICC, IF) as antibodies may perform differently across various techniques .

Many manufacturers conduct extensive testing during antibody development, excluding antibodies that display poor specificity or high cross-reactivity, but independent validation remains essential for critical research applications .

What methodology enables accurate ratio analysis in Cy3/Cy5 microarray experiments?

A validated approach for ratio analysis in Cy3/Cy5 microarray experiments involves reciprocal labeling of protein samples combined with a specialized calculation formula. As detailed in published literature, the methodology involves:

  • Dividing a protein sample into two aliquots and labeling one with Cy3 and the other with Cy5.

  • Preparing two microarray slides with reversed dye ratios:

    • Slide #1: Incubated with X amount of Cy3-labeled proteins and Y amount of Cy5-labeled proteins

    • Slide #2: Incubated with X amount of Cy5-labeled proteins and Y amount of Cy3-labeled proteins

  • Calculating the ratio for each target protein using the formula:
    R = [(Cy3Slide1 × Cy5Slide2)/(Cy5Slide1 × Cy3Slide2)]

Where Cy3Slide1 represents the background-subtracted mean intensity of Cy3 on slide #1 for a specific spot, and similarly for the other variables .

This approach normalizes for dye-specific effects and spot-to-spot variation, providing more reliable quantification. Validation experiments have confirmed that this methodology produces consistent ratio measurements across different protein targets when using controlled sample mixtures .

How can researchers optimize scanning parameters for Cy3/Cy5 microarrays?

Optimizing scanning parameters is essential for obtaining accurate and reproducible results:

By systematically optimizing these parameters, researchers can maximize signal-to-noise ratios while avoiding signal saturation or underdetection.

What approaches can detect and correct dye-specific biases in Cy3/Cy5 experiments?

Dye-specific biases can significantly impact experimental outcomes. Several strategies can identify and mitigate these effects:

  • Dye-swap experiments: Perform reciprocal labeling of samples with Cy3 and Cy5, then integrate results using specialized normalization algorithms .

  • Ratio calculation formula: Utilize the formula R = [(Cy3Slide1 × Cy5Slide2)/(Cy5Slide1 × Cy3Slide2)] from validated methodologies to normalize dye-specific effects .

  • Internal controls: Incorporate known reference proteins labeled with both dyes at defined ratios to establish normalization factors.

  • Statistical normalization: Apply computational methods such as LOWESS (Locally Weighted Scatterplot Smoothing) or quantile normalization to correct systematic dye biases.

  • Environmental factor control: Minimize exposure to factors known to differentially affect Cy3 and Cy5, such as ozone, which can disproportionately impact Cy5 stability .

These approaches, particularly when used in combination, can substantially reduce the impact of dye-specific biases on experimental outcomes.

How stable are Cy3 signals over time, and what storage conditions are recommended?

Research evidence demonstrates remarkable stability of Cy3 signals under appropriate storage conditions. A systematic study examined fluorescent intensities of Cy3 and Cy5 on antibody microarray slides preserved at -20°C and rescanned at days 10, 20, and 30 after the initial experiment . The results showed no significant changes in signal intensities compared to those scanned immediately after the standard microarray procedure .

For optimal signal preservation, the following storage conditions are recommended:

  • Store microarray slides in airtight slide vials

  • Maintain a constant temperature of -20°C

  • Allow slides to equilibrate to room temperature for approximately 30 minutes before rescanning

  • Protect from light exposure during handling

These findings provide practical guidance for researchers facing time constraints, indicating that microarray slides can be stored and scanned at a later time without compromising data integrity.

What environmental factors affect Cy3 signal intensity and how can they be controlled?

Several environmental factors can impact Cy3 signal intensity:

  • Light exposure: Cy3, like other fluorescent dyes, is susceptible to photobleaching. Minimize exposure to light during all experimental procedures and storage.

  • Ozone levels: While Cy3 appears less sensitive to ozone than Cy5, elevated ozone levels in laboratory atmospheres can potentially affect signal intensity. Consider using low-ozone environments for critical experiments .

  • Humidity: Excessive humidity may impact dye stability. Ensure microarray slides are thoroughly dried (e.g., by centrifugation at 1,000 × g) before scanning and storage .

  • Temperature fluctuations: Maintain consistent temperature during storage and handling to prevent condensation and potential signal degradation.

  • Chemical exposure: Avoid exposure to high pH solutions, alcohols, and other chemicals that may compromise dye stability.

Control strategies include working in climate-controlled environments, using ozone scrubbers when necessary, storing slides in sealed containers with desiccant, and minimizing handling time outside of controlled environments.

How can researchers address signal saturation issues in Cy3/Cy5 microarray experiments?

Signal saturation represents a significant challenge in microarray experiments. Several approaches can address this issue:

  • Preventative scanning optimization: Initially scan at reduced laser power and/or PMT settings to prevent saturation of high-abundance proteins.

  • Iterative scanning: If saturation is detected during data analysis, rescan the microarray slides with adjusted parameters. Research has confirmed that microarray slides can be stored at -20°C and rescanned later without significant signal degradation .

  • Extended dynamic range: Combine data from multiple scans with different sensitivity settings to extend the dynamic range of detection.

  • Sample dilution series: For highly abundant targets, prepare a dilution series to ensure measurements fall within the linear range of detection.

  • Protein concentration adjustment: Optimize the amount of protein loaded to avoid saturating high-abundance targets while maintaining detection of low-abundance proteins.

When implementing these approaches, maintain consistent scanning parameters across comparative samples to ensure valid data interpretation.

What strategies help resolve inconsistencies between Cy3 and Cy5 channels?

Channel inconsistencies can arise from various sources. Effective resolution strategies include:

  • Dye-swap normalization: Implement the reciprocal labeling approach described earlier, using the validated ratio calculation formula: R = [(Cy3Slide1 × Cy5Slide2)/(Cy5Slide1 × Cy3Slide2)] .

  • Balanced mixture experiments: Conduct control experiments with known protein mixtures at defined ratios (e.g., 2:1) to establish baseline channel performance .

  • Scanner calibration: Regularly calibrate your scanner using standardized fluorescent references to ensure consistent performance across channels.

  • Background subtraction optimization: Evaluate different background correction methods to identify the approach that minimizes channel-specific artifacts.

  • Data normalization algorithms: Apply computational methods specifically designed to address channel-specific biases in two-color microarray data.

By systematically implementing these strategies, researchers can significantly reduce inconsistencies between channels and improve data reliability.

How should researchers interpret conflicting results between antibody microarrays and other protein detection methods?

When faced with discrepancies between antibody microarray results and other methodologies:

  • Consider methodological differences: Antibody microarrays detect proteins in their native state, while techniques like Western blotting analyze denatured proteins. These fundamental differences may explain some discrepancies .

  • Evaluate dynamic range limitations: Different methods have varying dynamic ranges, potentially explaining quantitative inconsistencies for highly abundant or rare proteins.

  • Assess antibody performance: The same antibody may perform differently across various platforms. Verify antibody specificity in each application context.

  • Examine post-translational modifications: Some detection methods may be differentially sensitive to post-translational modifications that affect antibody binding.

  • Validate with orthogonal approaches: When critical discrepancies arise, implement a third independent method to resolve conflicts.

  • Consider sample preparation differences: Variations in protein extraction, labeling efficiency, or buffer composition may contribute to methodological discrepancies.

Researchers should view these techniques as complementary rather than contradictory, with each providing unique insights into protein expression and modification.

How do antibody microarrays using Cy3/Cy5 compare to other proteomics approaches?

Antibody microarrays offer distinct advantages and limitations compared to other proteomics methodologies:

Advantages:

  • Direct protein-level measurements, circumventing the poor correlation often observed between mRNA and protein levels

  • High-throughput capability, with platforms like ClontechTM allowing simultaneous examination of over 500 proteins

  • Relatively small sample requirements compared to some traditional proteomic techniques

  • Capacity to detect proteins across a wide range of cellular compartments (cytosolic, membrane-bound, and nuclear proteins)

Limitations:

  • Dependency on antibody availability and specificity

  • Potential cross-reactivity concerns

  • Limited dynamic range compared to some mass spectrometry approaches

  • Challenges in detecting post-translational modifications unless modification-specific antibodies are employed

Current applications focus on differential protein expression analyses across normal and pathological conditions, biomarker discovery, and pathway analysis . As antibody quality continues to improve and computational analysis methods advance, these platforms will likely become increasingly valuable complementary tools within the broader proteomics landscape.

What validation methods ensure reliable quantitative data from Cy3/Cy5 experiments?

Ensuring reliable quantification requires comprehensive validation:

  • Linearity assessment: Verify that signal intensities correlate linearly with protein abundance across the working range. High-quality antibody microarrays exclude antibodies that show non-linear binding kinetics .

  • Reciprocal labeling validation: Implement the dye-swap approach with the validated ratio calculation formula to normalize dye-specific effects .

  • Internal controls: Incorporate proteins at known concentrations and ratios to validate quantification accuracy across the microarray.

  • Technical replicates: Analyze duplicate or triplicate spots (as implemented in platforms like ClontechTM) to assess reproducibility .

  • Biological replicates: Perform independent biological replicates to distinguish technical variation from true biological differences.

  • Orthogonal validation: Confirm key findings using independent techniques such as Western blotting, ELISA, or targeted mass spectrometry.

  • Statistical validation: Apply appropriate statistical tests to determine the significance of observed differences, accounting for multiple testing correction.